Instructions for use

  • Do not use in automatic washers
  • Check materials compatibility before use

Manual process: Dose 2-20 ml of Manual Plus per liter of warm water (25-45°C).

  • Stir well.
  • Submerge instruments for 2-10 min.
  • Brush if necessary.

Ultrasonic process: Dose 2-20 ml of Manual Plus per liter of warm water (45-60°C).

  • Run ultrasonic bath for 2-10 min.
  • Thoroughly rinse instruments prior to disinfection.
  • Discard solution at least daily or after every 7-10 cycles.

Getinge Clean Manual Plus Detergent Instructions for Use-1Getinge Clean Manual Plus Detergent Instructions for Use-2Getinge Clean Manual Plus Detergent Instructions for Use-3

The chart is only a guide

Information on the reprocessing of medical devices from the manufacturer according to EN ISO 17664 must always be followed.

M in the diagram means: Always refer to instrument manufacturer’s guidelines.Getinge Clean Manual Plus Detergent Instructions for Use-4

Product information

  • Stabilized multi-enzyme formulation with powerful cleaning and sanitizing action
  • Actively sanitizes cleaning bath and maintains hygienic conditions up to 8 hours
  • Rapidly digests proteins, lipids, carbohydrates and biofilms
  • Non-corrosive, neutral pH, effective in hard and soft water
  • Fully biodegradable

Instructions for use

  1. Remove gross contamination from instruments.
  2. Measure the volume of the bath.
  3. Fill the bath with warm to hot water. 25-45°C preferred for manual cleaning and 45-60°C for ultrasonic baths.
  4. From the tables below determine the concentration of Getinge Clean Manual Plus required for the cleaning conditions applicable and add the required volume of the detergent to the pre- filled bath.

Manual cleaning

Getinge Clean Manual Plus Detergent Instructions for Use-5

Soiling levelWater qualityDos ageSoak time at 45 CSoak time at 25 C
LowGood2 mL/L3 min5 min
Poor5 mL/L4 min7 min
ModerateGood5 mL/L5 min8 min
Poor10 mL/L7 min10 min
HighGood10 mL/L7 min12 min
Poor20 mL/L10 min20 min

Ultrasonic bath cleaning

Getinge Clean Manual Plus Detergent Instructions for Use-6

Soiling levelWater qualityDos ageContact time at 50 CContact time at 25 C
LowGood2 mL/L1 min2 min
Poor6 mL/L1.5 min3 min
ModerateGood6 mL/L1 min2 min
Poor12 mL/L2 min4 min
HighGood15 mL/L3 min5 min
Poor20 mL/L4 min6 min